Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K071591

    Validate with FDA (Live)

    Date Cleared
    2008-04-16

    (310 days)

    Product Code
    Regulation Number
    866.3328
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RAMP® Influenza A/B Assay is a qualitative immunochromatographic assay used to identify the presence of Influenza A and Influenza B nucleoprotein antigens in nasal wash, nasal aspirate, nasopharyngeal aspirate, and nasopharyngeal swab specimens from symptomatic patients. It is an in vitro diagnostic assay that aids in the rapid differential diagnosis of influenza viral infections in symptomatic patients. A negative test is presumptive and it is recommended these results be confirmed by cell culture. Negative results do not preclude influenza virus infection and should not be used as the sole basis for treatment or other management decisions.

    The test performance characteristics for Influenza B were established primarily with retrospective, frozen specimens. Users may wish to further evaluate the sensitivity performance of this test for Influenza B using fresh samples.

    Device Description

    The RAMP Influenza A/B Assay is a qualitative immunochromatographic test that utilizes the RAMP 200 for the differential determination of Influenza B in nasal wash, nasal aspirate, nasopharyngeal aspirate, and nasopharyngeal swab samples. A wash/aspirate or swab sample is added to the Sample Buffer. The Sample Buffer is optimized to improve binding of the anti-influenza antibodies to the nucleoprotein antigens and reduce non-specific binding and fluorescent signal background. This sample is then mixed using the Assay Tip containing fluorescent-dyed particles conjugated to specific antibodies and applied into the sample well of the Test Cartridge. The sample migrates along the strip. Fluorescent-dyed particles coated with anti-Influenza A and anti-Influenza B antibodies bind to Influenza A or B antigens, respectively, if present in the sample. As the sample migrates along the strip, Influenza-bound particles are captured at either the Influenza A or the Influenza B detection zone, and excess fluorescent- dyed particles are captured at the internal standard zone.

    The instrument then measures the amount of fluorescence emitted by the complexes at the two detection zones (Influenza A and Influenza B) and at the internal standard zone. The instrument calculates a ratio (RAMP Ratio) using the fluorescence reading of each detection zone (A or B) and the internal standard zone. The instrument compares these ratios to pre-defined threshold limits to determine a positive or negative result for Influenza B in the tested sample.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly state pre-defined acceptance criteria in a dedicated section with target values. However, based on the performance study results, we can infer the observed performance metrics. The primary method of comparison is against cell culture.

    Metric (Inferred Acceptance Criteria based on study results)Reported Device Performance (RAMP® Influenza A/B Assay) - Prospective Fresh Clinical SpecimensNotes
    Influenza A
    Sensitivity (Nasopharyngeal Swab)80.3% (53/66) (95% CI: 68.7-89.1%)Ranges from 69.6% (age 22-59) to 91.7% (age 0-5) across subgroups.
    Specificity (Nasopharyngeal Swab)96.6% (541/560) (95% CI: 94.8-97.9%)Ranges from 94.0% (age 0-5) to 99.3% (age 22-59) across subgroups.
    Sensitivity (Nasal Wash/Aspirate)80.0% (36/45) (95% CI: 65.4-90.4%)Ranges from 0/0 (age 22-59 and >60, due to no positive cases) to 85.7% (age 0-5) across subgroups.
    Specificity (Nasal Wash/Aspirate)93.6% (162/173) (95% CI: 88.9-96.8%)Ranges from 89.1% (age 6-21) to 100% (age 22-59 and >60) across subgroups.
    Influenza B
    Sensitivity (Nasopharyngeal Swab)58.3% (7/12) (95% CI: 27.7-84.8%)Ranges from 50.0% (age 6-21, 22-59) to 100% (age 0-5) across subgroups. Lower sensitivity compared to Influenza A.
    Specificity (Nasopharyngeal Swab)98.0% (602/614) (95% CI: 96.6-99.0%)Ranges from 95.1% (age 6-21) to 99.4% (age 22-59, >60) across subgroups.
    Sensitivity (Nasal Wash/Aspirate)100% (9/9) (95% CI: 66.4-100%)Due to small numbers, some age groups showed 100% sensitivity but with wide confidence intervals or "NA". Overall 100% sensitivity reported for this collection type.
    Specificity (Nasal Wash/Aspirate)99.0% (207/209) (95% CI: 96.6-99.9%)Ranges from 98.3% (age 6-21) to 100% (age 22-59 and >60, due to no positive cases).
    Analytical Precision (Agreement with Expected Results)100% (540/540) for Influenza A (High Negative, LoD, 2x LoD positive) and Influenza B (High Negative, LoD, 2x LoD positive) samples across multiple operators, sites, and days.The overall RAMP Ratio %CV ranged from 9.6% to 15.1%. This demonstrates excellent reproducibility and reliability for quantitative measurements, which supports the qualitative output.
    Analytical Sensitivity (LoD)Influenza A: 3.0x10² to 6.4x10² EID50/mL (VTM/PBS) Influenza B: 2.8x10² to 7.1x10⁴ EID50/mL (VTM/PBS)These indicate the lowest concentration of virus the assay can reliably detect.
    Analytical Reactivity100% Positive for all tested strains of Influenza A (5 strains) and Influenza B (3 strains) at specified concentrations.Confirms the device's ability to detect various common strains of Influenza A and B.
    Analytical Specificity (Cross-Reactivity)Negative for all 15 viruses and 17 bacteria tested.Shows the device does not cross-react with other common respiratory pathogens or microbes found in the nasal cavity. Note: Chlamydophilia pneumoniae not determined.
    Interference100% agreement with expected results (negative for negative samples, positive for positive samples) in the presence of 23 potentially interfering substances (medications, OTC products, whole blood up to 2% v/v).Demonstrates robustness against common interfering substances. A warning is included for visibly bloody samples.
    Transport Media Compatibility100% Agreement (Negative for negative, Positive for positive) across 6 commercially available and 2 clinical-site-prepared transport media.Ensures the device performs reliably with various transport media.
    Sample Collection Swab Compatibility100% Negative for negative swab samples. For LoD positive samples, agreement ranged from 67% to 100% positive for Influenza A and 67% to 100% positive for Influenza B, depending on swab material.Polyester and Rayon swabs showed 67% detection at LoD for Influenza A and B respectively (compared to 100% for Foam and Nylon). Calcium alginate swabs are noted as not recommended. This indicates slight variability based on swab material, particularly at lower viral loads.

    2. Sample Size and Data Provenance:

    • Test Set (Clinical Performance):

      • Prospective Study: 844 fresh specimens (nasal wash/aspirate, nasopharyngeal aspirate, or nasopharyngeal swab samples). This study was conducted in North America during the 2006-2007 influenza season.
      • Retrospective Study: 75 retrospective frozen clinical nasopharyngeal swab samples and 130 retrospective frozen clinical wash/aspirate samples. These were compared to original cell culture results performed on fresh specimens.
    • Training Set: The document does not explicitly mention a training set size or methodology for the device's algorithm. The studies described are for analytical (device characteristics) and clinical performance evaluation, not for training a machine learning model. This is a traditional IVD.

    3. Number of Experts and Qualifications for Ground Truth (Test Set):

    • The ground truth for the clinical performance studies was established using cell culture. The document does not specify the number or qualifications of experts involved in performing or interpreting the cell culture results. It states that "cell culture testing was performed using fresh specimens" and implies these are standard laboratory procedures.

    4. Adjudication Method (Test Set):

    • The document describes comparison studies where the RAMP® Influenza A/B Assay results were compared directly against cell culture results. The term "adjudication method" typically applies to resolving discrepancies between multiple readers or between a new device and a reference standard where a third, independent assessment is needed. In this case, cell culture acts as the definitive gold standard, and direct comparison is made. There is no mention of a separate adjudication process beyond the cell culture determination.

    5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. This device is an in-vitro diagnostic (IVD) assay that produces a direct result, not an AI-assisted diagnostic imaging tool that improves human reader performance. The "readers" here are the laboratory personnel performing the assay.

    6. Standalone Performance Study:

    • Yes, a standalone performance study was done for the device. The "Clinical Performance" section directly compares the RAMP® Influenza A/B Assay (algorithm only, as it generates the result) against the cell culture gold standard. The sensitivity and specificity results presented in the tables are indicative of the algorithm's standalone performance.

    7. Type of Ground Truth Used:

    • For the clinical performance studies, the ground truth was viral cell culture. This is explicitly stated: "The performance of the RAMP Influenza A/B Assay was compared to cell culture."

    8. Sample Size for the Training Set:

    • As mentioned in point 2, the document does not describe the training of a machine learning algorithm. Therefore, there is no explicit training set size provided for this device in the context of AI/ML. The analytical studies describe the development and validation of the assay's operational parameters rather than training a model.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no mention of an AI/ML training set in the document. For the development of the assay (which could be considered analogous to "training" in a traditional IVD context), the process involves determining optimal reagent concentrations, binding efficiencies, and establishing cut-off values. This is based on internal analytical studies (like LoD, precision, reactivity) using characterized viral strains and clinical samples, rather than a separate ground truth establishment for a training set in the AI sense.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1